The Transducer Block
3.6 Transducer Block Examples Setup
SITRANS TH400 PROFIBUS® PA
22
Configuration Manual, 02/2007, A5E01039143-01
Example:
The coordinates for converting the signal from a logarithmic potentiometer to a linear signal.
TAB_ACTUAL_NUMBER = 10 (number of linearisation points to follow up to max 50)
TAB_XY_VALUE1
= 0,0; -100
TAB_XY_VALUE2
= 0,1; 0
TAB_XY_VALUE3
= 0,2; 100
TAB_XY_VALUE4
= 0,4; 200
TAB_XY_VALUE5
= 0,8; 300
TAB_XY_VALUE6
= 1,6; 400
TAB_XY_VALUE7
= 3,2; 500
TAB_XY_VALUE8
= 6,4; 600
TAB_XY_VALUE9
= 12,8; 700
TAB_XY_VALUE10
= 25,6; 800
(Output will readout 325% with 1.0% potentiometer value)