Modbus holding registers
A.3 Modbus holding registers tables
FC410 with Modbus
Operating Instructions, 02/2016, A5E33120874-AC
129
Table A- 10 Device diagnostics
Modbus
address
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value range Access
level
2756
Float / 4
Driver Current
Actual sensor driver current.
The actual driver current is
viscosity and sensor size
dependent
- [A]
0 to 0.124
Read Only
2758
Float / 4
Pick-up Amplitude 1 Actual pick-up 1 amplitude
- [V]
0 to 0 9999 Read Only
2760
Float / 4
Pick-up Amplitude 2 Actual pick-up 2 amplitude
- [V]
0 to 0 9999 Read Only
2762
Float / 4
Sensor Frequency
Actual sensor frequency
- [Hz]
0 to 1 023
Read Only
3032
Float / 4
PCB Temperature
Actual sensor electronic
temperature
- [C°]
-50 to 200 Read Only
Table A- 11 Aerated flow
Modbus
address
Data type /
Size [bytes]
Parameter
Description
Default value
[unit]
Value range Access
level
2200
Unsigned /
2
Aerated Flow Alarm
Limit
Alarm limit calculated in per
cent of accepted bad
measurements.
80 [%]
0 to 99
Read /
Write
2201
Unsigned /
2
Aerated Flow Warn-
ing Limit
Warning limit calculated in
per cent of accepted bad
measurements
0 [%]
0 to 99
Read /
Write
2202
Unsigned /
2
Measurement Samp-
le Time
The time period over which
the actual percentage of
bad measurements is
calculated
5 [s]
1 to 10
Read /
Write
2203
Unsigned /
2
Aerated Flow Filter
Aerated flow filter
0: Disabled
1: Enabled
2: Auto
Auto means that filtering
starts automatically when
aerated flow is measured.
2
0 to 2
Read /
Write
2204
Unsigned /
2
Filter Time Constant PV Filter Time Constant
0: 0.5 seconds
1: 1 second
2: 2 seconds
3: 5 seconds
4: 10 seconds
5: 20 seconds
6: 30 seconds
7: User Defined Value
4
0 to 7
Read /
Write
2205
Float / 4
Filter Start Hystere-
sis
The filter is active when the
hysteresis value is
exceeded.
Aerated Flow Filter must be
set to Auto
0.015 [V]
0 to 0.124
Read /
Write