8. Commissioning
Commissioning
SFIDK.PS.029.V3.02 -
521H0553
SITRANS F US SONOFLO
®
®
®
®
®
Factory settings
On start-up the meter uses the factory default setting in the SENSORPROM
®
memory unit. The
following tables state the factory-set values. The range for each setting is given.
If a setting's range is exceeded, the cursor moves to the first character in the display and
flashes to indicate that the setting is invalid. The selected setting cannot be locked with
the LOCK key until a valid figure is selected. If the menu is left with cursor flashing, the
original value is maintained.
Since a number of factory-set values, e.g. max. volume flow, max. mass flow etc., are dimension-
dependent, these are given separately in the table "Dimension-dependent settings".
Table 1
Factory settings
Settings available
BASIC SETTINGS
Flow direction definition
Pos
Pos, Neg
Volume flow unit
m
3
/h
Volume flow max.
1300
Sound velocity, min.
1400
400 - 2000
Sound velocity, max.
1600
400 - 2000
Mass flow
Off
Scaled, O
2
, N
2
, Ar, Off
Density unit
kg/m
3
g/cm
3,
kg/m
3
, ton/m
3
, lb/ft
3
Sound Velocity 1
1447
400 - 2000
Density 1
997.7
Sound Velocity 2
1542
400 - 2000
Density 2
998.0
Mass flow unit
ton/h
x x
Mass flow max.
1300
Totalizer 1
Off
Volume flow, Mass flow, Off
Totalizer 1 direction
Pos
Pos, Neg, Bidirectional
Totalizer 1 unit
m
3
Totalizer 2
Off
Volume flow, Mass flow, Off
Totalizer 2 direction
Pos
Pos, Neg, Bidirectional
Totalizer 2 unit
m
3
Low flow cut-off
1.5%
0 - 9.9%
Force control
Off
Off, Max., Min.
OUTPUT SETUP
Current 1
Off
Volume flow, Mass flow, Sound Velocity, Off
Current 1 direction
Unidirectional
Unidirectional, Bidirectional
Current 1 range
4 - 20 mA
0 - 20 mA, 4 - 20 mA
Current 1 Time constant
5 s
0.8 - 30 s
Current 2
Off
Volume flow, Mass flow, Sound Velocity, Off
Current 2 direction
Unidirectional
Unidirectional, Bidirectional
Current 2 range
4 - 20 mA
0 - 20 mA, 4 - 20 mA
Current 2 Time constant
5 s
0.8 - 30 s
Current 3
Off
Volume flow, Mass flow, Sound Velocity, Off
Current 3 direction
Unidirectional
Unidirectional, Bidirectional
Current 3 range
4 - 20 mA
0 - 20 mA, 4 - 20 mA
Current 3 Time constant
5 s
0.8 - 30 s
Frq.1/Pulse 1
Off
Volume flow, Mass flow, Sound Velocity, Off
Frq. 1/Pulse 1 direction
Unidirectional
Unidirectional, Bidirectional
Frq. 1/Pulse 1
Pulse
Frequency, Pulse
Pulse 1 unit
m
3
L, L
×
10, L
×
100, m
3
, m
3
×
10, m
3
×
100,
UG
×
0.1, UG, UG
×
10, UG
×
100, MUG,
kg, kg
×
10, kg
×
100, ton, lb, lb
×
10, lb
×
100
Pulse 1 Volume/pulse
1
Pulse 1 Width
50 ms
50
µ
s, 500
µ
s, 50 ms E. Mech., 500 ms, 1 s, 5 s
Frq. 1 frequency
10 kHz
500 Hz, 1 kHz, 5 kHz, 10 kHz
Frq. 1 time constant
5 s
0.8 - 30 s