46
8. Commissioning
Commissioning
SFIDK.PS.029.V3.02 -
521H0553
SITRANS F US SONOFLO
®
®
®
®
®
PIPE GEOMETRY
In the PIPE GEOMETRY menu the number of tracks is defined. Here the signal converter can be
set to a 1-4 track system. If the signal converter is connected to a sensor with a SENSORPROM
®
memory unit the data in the PIPE GEOMETRY menu are read-only. If the signal converter is used
with a sensor without a SENSORPROM
®
unit, e.g. with a type SONO 3110 mounting kit the pipe
data must be entered here. The standard configuration of a SONO 3000 is a two-track system. It
can easily be used as one-track system or in special cases as a three or four track system.
Using the SONO 3000 as a three or four track system is beyond the scope of this manual.
Please contact Siemens Flow Instruments for further information.
There is a connection between the number of tracks and which tracks may be set. If installed as
a one-track meter, the meter uses only the values set in track 1. Furthermore, it has only electrical
connections to transducers C and D. If the meter is installed as a two track meter, one of the tracks
can be switched off by moving to the required track and changing the setting to "OFF".
PIPE
GEOMETRY
UNIT
PIPE
DIAMETER
ROUGHNESS
NO. OF
TRACKS
TRACK 1
TRACK 2
TRACK
ON/OFF
TRACK
ON/OFF
LENGTH
LENGTH
ANGLE
ANGLE
DISPLACEMENT
DISPLACEMENT