34
7. Electrical connections
Electrical connections
SFIDK.PS.029.V3.02 -
521H0553
SITRANS F US SONOFLO
®
®
®
®
®
Frequency output with a
load > 10 k
Ω
Ω
Ω
Ω
Ω
If the load exceeds 10 k
Ω
it is recommended that a resistor is connected to the frequency output
as shown in the figure above.
Relay mode signal con-
verter, compact version
Relay shown in de-energized position.
Relay 1
OFF
ON
Connection between
(De-energized)
(Energized)
terminal no.
45-46
44-46
Error
“Error”
"Normal"
Direction
“Forward”
"Reverse"
Sound limit
Inside range
Outside range
Relay mode signal
converter, 19" insert
Relays shown in de-energized position.
Relay 1
OFF
ON
Connection between
(De-energized)
(Energized)
terminal no.
44-45
45-46
Error
“Error”
"Normal"
Direction
"Forward"
"Reverse"
Sound limit
Inside range
Outside range
Relay 2
OFF
ON
Connection between
(De-energized)
(Energized)
terminal no.
41-42
42-43
Direction
"Forward"
"Reverse"
Sound limit
Inside range
Outside range
Relay 2
Relay 1