6. Installation of the signal converter
Installation
SFIDK.PS.029.V3.02 -
521H0553
SITRANS F US SONOFLO
®
®
®
®
®
Remote installation of the
signal converter, IP 67
version
Add approx. 30 mm to each side to allow for cables.
Wall or panel mounting
Installation in horizontal pipe
Installation in vertical pipe
1. Use a 75
Ω
coaxial cable between feeler and
remote signal converter.
2. Mount the wall bracket.
3. Remove the connection board and loosen
the earth cable.
4. Connect track 1 transducer to terminals
85...88 and track 2 transducer to terminals
81...84.
Signal wires to even, screen to odd numbers.
5. Reinstall the earth cable and snap in the
connection board.
6. Connect current and signal cables and tighten
all cable entries to ensure optimum sealing.
7. Mount the SENSORPROM
®
unit supplied
with the set on the connection board.
8. Place the signal converter in the wall bracket.