37
Commissioning
8. Commissioning
SFIDK.PS.029.V3.02 -
521H0553
SITRANS F US SONOFLO
®
®
®
®
®
The menu structure of the SONO 3000 signal converter is shown in an overview map and in detail
in the description of the layout of the submenus .
The menu is divided in two parts. An OPERATOR MENU and a SETUP MENU. The SETUP MENU
is accessed by pressing the TOP UP key
for 2 seconds. The SETUP MENU operates in two
modes: VIEW and CHANGE. VIEW is a read-only mode and CHANGE is both a read and write mode.
The preselected flowmeter settings can only be read in VIEW mode. The access to CHANGE mode
is protected with a USER CODE. The user code is factory-set at 1000.
The signal converter always starts in the basic OPERATOR MENU showing the flow rate.
The PAGE FORWARD and PAGE BACKWARD keys
are used to step through the
OPERATOR MENU.
The SETUP MENU contains the following submenus:
•
CONVERTER SETUP MODE
•
RESET MODE
•
SERVICE MODE
•
LANGUAGE SETUP
•
OPERATOR MENU SETUP
•
USER CODE SETUP
The PAGE FORWARD key
is used to step through the main menu. Pressing the LOCK key
opens the submenu below. There is no access from the submenu back to the SETUP main menu.
To leave the submenu press the TOP UP key
, which returns the program to the OPERATOR
MENU. To perform further changes in other main menus, press the TOP UP key
for 2 seconds,
then press the PAGE FORWARD key
to select the CHANGE mode. Enter the USER CODE,
press the LOCK key
and page through the main menu using the PAGE FORWARD key
until the required menu is reached. Access the submenu by pressing the LOCK key
.
CONVERTER SETUP MODE contains 3 submenus:
•
BASIC SETTINGS
•
OUTPUT SETUP
•
SENSOR CHARACTERISTICS
In BASIC SETTINGS selection of flow direction, measuring range, measuring units, totalizer units
and low flow cut off can be made.
In OUTPUT SETUP the required output signals can be selected.
SENSOR CHARACTERISTICS gives sensor information. This information is uploaded automatically
from the SENSORPROM™ memory unit.
In the case of remote installation, when the signal converter is connected to the sensor by means
of a signal cable, the length of the cable must be entered into APPLICATION PARAMETERS, a
submenu of SENSOR CHARACTERISTICS
In RESET MODE totalizers are reset, zero adjustmen is performed and the menus can be reset to
factory settings.
When setting up using a 9 V battery, totalizers are not reset.
NOTE!
USER CODE, CORRECTION FACTOR, LANGUAGE SETUP and AUTO ZERO do not return to the
factory setting via RESET MODE.
SERVICE MODE provides the facility to set the outputs to fixed values (forced outputs).
On leaving SERVICE MODE, all settings made in SERVICE MODE are cancelled.
The menu language can be selected.
The information accessible to the operator can be selected in this menu.
Menu structure
OPERATOR MENU
CONVERTER SETUP MODE
BASIC SETTINGS
RESET MODE
OUTPUT SETUP
SENSOR
CHARACTERISTICS
SETUP MENU
LANGUAGE SETUP
SERVICE MODE
OPERATOR MENU SETUP