25
Installation
6. Installation of the signal converter
SFIDK.PS.029.V3.02 -
521H0553
SITRANS F US SONOFLO
®
®
®
®
®
Signal converter
IP 67 version
Setting the supply voltage
The mains voltage selector 115/230 V a.c. is
located at the bottom of the signal converter.
1
2
4
3
5
Setting prior to installation
The flowmeter can be set before the final in-
stallation using a 9 V alkaline battery. This is
especially advantageous if the flowmeter is to
be installed before the system is put into oper-
ation.
1. Position the SENSORPROM
®
memory
unit supplied in the terminal box at the
bottom of signal converter as shown.
2. To protect the pins, place the signal con-
verter in the packaging base supplied.
3. Set the signal converter as described un-
der “Commissioning”.
The SENSORPROM
®
unit has now been
programmed with the required settings.
4. Remove the battery and replace the SEN-
SORPROM
®
unit in the terminal box. The
flowmeter is now ready for operation.
Turning the control pad
1. Remove the outer frame using a fingernail
or a screw driver.
2. Loosen the 4 screws retaining the control
pad.
3. Withdraw the control pad and turn it to the
required position.
4. Tighten the 4 screws until a mechanical
stop is felt in order to obtain IP 67 enclo-
sure rating.
5. Snap-lock the outer frame onto the control
pad (click).