SFIDK.PS.029.Q3.02
39
7. Starting up
Starting up
SITRANS F US SONOFLO
®
®
®
®
®
7.4 Sub menus
The figure on previous page gives a view of the menu build up. Below is a more detailed description
of the submenus. Use menu block diagram to step through to required sub-menu. (See also detailed
diagram blocks).
Information regarding the service menu is placed in the section "Trouble shooting".
The flow direction is defined for volume and mass flow simultaneously.
Flow direction
Volume flow
User code setup
The user code can be changed in this menu.
The code is factory-set at 1000. If the user code is lost, the factory setting can be re-established as
follows: Switch off supply voltage, press the TOP-UP key
and switch on the supply voltage.
Release the key after ROM and RAM tests are completed. The user code is now reset to 1000.
If the signal converter is left in CONVERTER SETUP for more than 10 minutes, the converter
automatically reverts to OPERATOR MENU.
VOL. FLOW
MAX.
VOLUME
FLOW
VOLUME FLOW
UNIT: m
3
/s
UNIT: m
3
/min
UNIT: m
3
/h
UNIT: l/s
UNIT: l/min
UNIT: l/h
UNIT: ml/s
UNIT: UGPS
UNIT: UGPM
UNIT: UGPH
UNIT: MUGPO
UNIT: Ml/d
MAX RANGE
Choose engineering unit and the maximum volume flow range.
For SONO 3000 not fitted with a SENSORPROM
®
unit the display indication must be formatted.
It is done as follows:
a) Sensor
≤
DN 2000:
In the menu "MAX. RANGE" a value 10 x "VOL. FLOW MAX." should be entered.
b) Sensor
>
DN 2000:
In the menu "MAX. RANGE" a value 5 x "VOL. FLOW MAX." should be entered.
(Max. 540,000 m³/h).
Sound velocity
The engineering units for the sound velocity are in meters per second. The measuring range is set
by entering the minimum and the maximum sound velocity.
Error relay and error indication are activated, if the sound speed is outside the limits stated.
SOUND
VELOCITY
SOUND
VELOCITY
MIN.
SOUND
VELOCITY
MAX.