5. Signal converter installation
Signal conv. inst.
SITRANS F US SONOFLO
®
®
®
®
®
The remote version is installed as shown in the figure. The electronics and the display/keypad unit
can be positioned in the same way as those for the compact version.
5.12.7 Remote installation
of signal converter
5.12.8 Electrical connec-
tion of separate
version
Electrical connections are made in the terminal housing located on the wall fitting.
No connections
are made in the front of the signal converter
.
The terminal housing is divided so that all terminals in the "e" increased safety category are located
on the left-hand side. This comprises voltage supply and sensor connections. All the output
terminals located on the right hand side are intrinsically safe in category "ia".
The cap over the power supply terminals must be removed
before cable installation