22
SFIDK.PS.029.Q3.02
Signal conv. inst.
5. Signal converter installation
SITRANS F US SONOFLO
®
®
®
®
®
5.6
19" insert
The 19" insert is designed for installation in a 19" rack. The insert has a width of 28 TE (142 mm),
a height of 3 U (128 mm) and a module depth of 160 mm.
The insert can be installed directly in a 19" rack or on a wall in the IP 65 version. An optional mounting
kit for
front of panel
mounting and
back of panel
mounting is available. The various types of
installation are shown on the following pages.
5.6.1 Installation in 19"
rack
1. Fit the SENSORPROM
®
unit on the "mother board" (The SENSORPROM
®
unit is located in the
terminal box of the SONO 3300 sensor).
2. Mount the "mother board" and the guide rails in the rack system as shown.
3. Connect the cables as shown under "Electrical connection".
4. Plug the signal converter into the rack system.
5.6.2 Signal converter in
IP 65 version
1. Mount the IP 65 housing to the wall with four screws.
2. Remove the signal converter from the IP 65 housing and mount the SENSORPROM
®
unit on
the "mother board" located in the back of the housing.
(The SENSORPROM
®
unit is located in the terminal box of the sensor).
3. Connect the cables to the terminals accessible in the connection box located separately under
the housing, see "Electrical connection".
4. Plug in the signal converter and close the cover.