20
SFIDK.PS.029.Q3.02
Signal conv. inst.
5. Signal converter installation
SITRANS F US SONOFLO
®
®
®
®
®
1. Release the earth lead from the bottom of the terminal box.
Remove and keep all the leads connecting the terminal box to the plate carrying the terminal
boards.
Move the plate to the wall bracket and fit the earth lead in the base of the wall bracket.
There are two types of wall brackets. One for a twisted cable pair and one for 4 coaxial cables.
Terminal box
Terminal box
Wall bracket
Terminal box
Wall bracket
SONO 3300/3000 remote IP 67
2.
Wall bracket for mounting of multi coax cable:
Always use the cable entry nearest the
associated terminal so that unscreened lead ends from one cable do not cross those from the
other cable.
5.4 Wall bracket for use
with multicoaxial
cable
5.3 Remote installation of
the signal converter
With
remote
installation, the distance between
sensor and signal converter must not exceed 20
m when using multi-coaxial cable. Is a longer
distance required use of coaxial cable is
recommended.