7. Starting up
Starting up
SITRANS F US SONOFLO
®
®
®
®
®
The menu structure of the SONO 3000 signal converter is shown in an overview map and in detail
by the build-up of each submenu.
The menu is built up in two parts. An OPERATOR MENU and a SETUP MENU. Access to the
SETUP MENU is, by pressing the TOP-UP key
for 2 seconds. The SETUP MENU will operate
in two modes: VIEW and CHANGE mode. VIEW is a read only mode and CHANGE is both a read
and write mode. The preselected flowmeter settings can only be scanned via VIEW MODE. The
access to CHANGE MODE is protected with a USER CODE. The user code is factory-set at 1000.
The signal converter always starts in the basic OPERATOR MENU showing flow rate.
The PAGE FORWARD and PAGE BACKWARD key
is used to step through the
OPERATOR MENU.
The SETUP MENU contains the following menues:
•
Converter set-up mode
•
Reset mode
•
Service mode
•
Language set-up
•
Operator menu set-up
•
User code set-up
The PAGE FORWARD
key is used to step through the main menu. Pressing the LOCK key
opens the submenu below. There is no access from the submenu back to the main menu. To leave
the submenu press the TOP-UP key
, which returns the program to the OPERATOR MENU.
To perform further changes in other main menus, press the TOP-UP key
for 2 seconds, then
press the PAGE FORWARD key
to select the CHANGE mode. Enter the USER CODE, press
the LOCK key
and page through the main menu using by the PAGE FORWARD key
until
the required menu is reached. Access the submenu by pressing the LOCK key
.
CONVERTER SETUP MODE contains 3 submenus:
•
BASIC SETTINGS
•
OUTPUT SETUP
•
SENSOR CHARACTERISTICS
In BASIC SETTINGS selection of flow direction, measuring range, measuring units, totalizer units
and low flow cut-off can be made. Further, overriding of the outputs in an error situation can be
established.
In OUTPUT SETUP the required output signals can be entered.
SENSOR CHARACTERISTICS gives sensor information. This information is uploaded automatically
from the SENSORPROM
®
unit.
With remote installation, e.g. when the signal converter is connected to the sensor by means of a
signal cable, it is necessary to enter the length of this cable into the meter. The cable length is
entered in the SENSOR CHARACTERISTICS menu under APPLICATION PARAMETERS.
In RESET MODE the totalizers can be reset, the flowmeter zero point adjusted and the factory
setting re-established.
NOTE!
USER CODE, CORRECTION FACTOR, LANGUAGE and AUTO ZERO do not return to the factory
setting via RESET MODE.
SERVICE MODE provides the facility to set the outputs to fixed values (forced outputs). On leaving
SERVICE MODE, all settings, made in SERVICE MODE, are cancelled.
The menu language can be selected.
The information accessible to the operator can be selected in this menu.
7.2 Menu build-up
Basic settings
Output setup
Sensor characteristics
Language setup
Service mode
Reset mode
Converter setup mode
Setup menu
Operator menu
Operator menu setup