5. Signal converter installation
Signal conv. inst.
SITRANS F US SONOFLO
®
®
®
®
®
5.2 Installation of the
signal converter in
compact mode
With
compact
installation the temperature of
the liquid must not 120
°
C and the
ambient temperature must not 50
°
C.
The signal converter in compact version can be
installed indoors or outdoors.
The ambient temperature must not exceed –20
and 50
°
C. The signal converter must not be ex-
posed to direct sunlight. Use a shield if neces-
sary.
Terminal box
1. Remove and keep terminal box lid.
2. Turning the terminal box.
Lift and move the plate with terminal blocks to
one side to give access to the screws. Re-
move the two diametrically opposed screws
and loosen the remaining two. Turn the ter-
minal box to the required position and retighten
the screws.
Warning:
If all four screws are removed at the same
time there is a risk of damaging the leads to
the sensor.
3. Connect cables as shown under "Electrical
connection".
4. Tighten the cable glands to obtain optimum
sealing. The gaskets must protrude along the
cable.
5. Plug the signal converter into the terminal
box.
T
max
. = 120 °C