NOTICE
HART communication
HART communication on active current output could be disturbed during operation in areas
with a high frequency electromagnetic field
NOTICE
Ouptput cables
If long cables are used in noisy environments, it is recommended to use screened cables.
NOTICE
Electrode cables
Dotted connections are only to be made when using special electrode cables.
Active/passive current output
83N19.11
Pos
Active/Passive
Factory default
A
Active
MAG 6000 I
B
Passive
MAG 6000 I Ex version
Note
For Ex versions active or passive current output is preselected at ordering and cannot be
changed.
Non-Ex versions can be connected as either active or passive.
Note
The factory default setting for the current output is marked by an "X" and can be found on a
label on the back of the lid.
Connecting
5.2 Electrical connection
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018,
A5E02083319-AH