Non-Ex version
Ex version
Not connected
Shield for BUS Comm.
97
Fieldbus
Foundation
Profibus PA
95
96
1
2
Modbus RTU
T2
94
T1
91
92
93
B
A
Profibus DP
77
78
Digital input
Ri 4.4 Kohm
-
+
-
+
200 Ohm<Rl<10 Kohm
Max. 100mA
DC 3-30V
Digital output
57
56
DC 24V / 1A
AC 42V / 2A
Relay output
COM
NC
NO
46
45
44
HART
Current output
-
+
Gnd
Ground
32
31
N/-
N (2)
AC/DC
L/+
L (1)
Description:
Function:
Terminals:
RL<500 Ohm, HART >230 Ohm
DC 11-30V
Electrical Connection
I-OUT ACTIVE
I-OUT PASSIVE
X indicate factory default setting
Modul addr.:
FW revision:
HW revision:
Not connected
Shield for BUS Comm.
97
Fieldbus
Foundation
Profibus PA
95
96
Ui=17.5V, Ii=380mA
Pi=5.32W, Ci=0
Li=0
1
2
Modbus RTU
T2
94
T1
91
92
93
B
A
Profibus DP
77
78
Digital input
Pi=1.2W, Ci=0
Li=0
-
+
-
+
200 Ohm<Rl<10 Kohm
Pi=1.2W, Ci=11nF,
Ui=28V, Li=100mA
Digital output
57
56
Li=0
Pi=1.2W, Ci=7.5nF
Ui=30V, Ii=200mA
Relay output
COM
NC
NO
46
45
44
HART
Current output
-
+
Gnd
Ground
32
31
N/-
N (2)
Um=264V
AC/DC
L/+
L (1)
Description:
Function:
Terminals:
Passive: Ui=30V/DC, li=100mA
Active: Uo=28V, lo=87mA
RL<500 Ohm, HART >230 Ohm
Ui=30V,
I-OUT ACTIVE
I-OUT PASSIVE
X indicate factory default setting
Modul addr.:
Electrical Connection
Approval: 11ATEX2124X
Li=34 µH
FW revision:
HW revision:
①
Fill in the address of the used communication module after installation
②
Fill in the new FW and HW revisions in case spare part exchange
Note
Min. 4 mm
2
is only required for ATEX 2 GD and FM/CSA Class I, Zone 1/21 installations.
Note
Terminals 81 and 84 are only to be connected if special electrode cable with double screening
is used, e.g. when empty pipe function or long cables are used.
Note
Special cable with individual wire shields (shown as dotted lines) are only required when using
empty pipe function or long cables.
Note
All inputs and outputs are galvanically isoloated PELV circuits with 60 VDC isolation from each
other and ground. Exception are sensor connections that are connected to ground.
WARNING
Grounding
The mains protective earth wire must be connected to the PE terminal in accordance with the
diagram (Protection class I).
Connecting
5.2 Electrical connection
32
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018,
A5E02083319-AH