Compact transmitter version:
● Ex d e [ia] ia IIC T3-T6 Gb
Ex tD A21 IP67 T** °C (** pipe tempe13K, but not less than T85 °C
(MAG 3100 / MAG 3100 P)
● Ex d e [ia] ia IIB T3-T6 Gb
Ex tD A21 IP67 T** °C (** pipe tempe18K, but not less than T85 °C
(MAG 1100 / MAG 1100 F)
*
Y
HU
BQ
Q
For sensors:
Refer to sensor certificates:
MAG 3100 / MAG 3100 P: Sira 07ATEX1182X (DN 15-DN300); Sira 03ATEX3339X (DN350-
DN2000)
MAG 1100 / MAG 1100 F: Sira 07ATEX3181X
Safety parameters associated with the remote transmitter version
Table 2-1
Intrinsically safe data
Sensor electrode input (Terminals 0, 81, 82, 83, 84 - "ia circuits")
IIB
IIC
U
O
30 V DC
30 V DC
I
O
6.1 mA
6.1 mA
C
O
560 nF
66 nF
L
O
1 H
0.96 H
P
O
45.5 mW
45.5 mW
Safety notes
2.3 Installation in hazardous area
14
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018,
A5E02083319-AH