SITRANS F M MAGFLO
®
®
®
®
®
24
SFIDK.PS.026.E5.02
ENGLISH
Communication mode
Basic settings
Operator active
Service mode
Output
Operator inactive
Operator menu
External input
Product identity
Sensor characteristics
Language mode
Reset mode
5.
Commissioning
5.1 Keypad and display
layout
Keypad
The keypad is used to set the flowmeter. The function of the keys is as follows:
TOP UP KEY
This key (hold 2 sec.) is used to switch between operator menu
and setup menu. In the transmitter setup menu, a short press will
cause a return to the previous menu.
FORWARD KEY
This key is used to step forward through the menus. It is the only
key normally used by the operator.
BACKWARD KEY
This key is used to step backward through the menus.
CHANGE KEY
This key changes the settings or numerical values.
SELECT KEY
This key selects the figures to be changed.
LOCK/UNLOCK KEY
This key allows the operator to change settings and gives
access to submenus.
Display
The display is alphanumerical and indicates flow values, flowmeter settings and error messages.
The upper line is for primary flow readings and will always show either flow rate, totalizer 1 or totalizer
2. The line is divided into 3 fields.
S:
Sign field
P:
Primary field for numerical value
U: Unit field
The centre line is the title line (T) with individual information according to the selected operator
or setup menu.
The lowest line is the subtitle line (ST) which either will add information to the title line or keep
individual information independent of the title line.
F:
The alarm field. Two flashing triangles will appear by a fault condition.
M: The mode field. The symbols indicate the following.
Ready for change
Access to submenu
Value locked
RESET MODE: Zero setting of
totalizers and initialization of setting
L:
The lock field. Indicates the function of the lock key.
5.
Commissioning
With the capacitive touch keypad operation is achived without any open lids. An LED light gives
a feedback.