Diagnostic Screens
for Path-2.
2. Refer to the Data Capture file heading list that provides the variable name and a short
description of each column within the report. For convenience, the short variable name is
provided above the first row of data.
3. The "summary" rows found at the bottom of the file include the following information for
applicable data items:
Summary Item
Description
Average Value
Average of all data within column.
Minimum
Minimum value within column.
Maximum
Maximum value within column.
Warning Limit
Warning limit used for Diagnostics Screen "Warning"
display.
Failure Limit
Failure limit used for Diagnostics Screen "Fail" display.
Number of Warnings
Number of Warnings occurring within the capture period for
each data item.
Number of Failures
Number of failures occurring within the capture period for
each data item.
Number of Samples
the number of Data reports collected in the Data
Capture file
Table 5-1 Data Capture Summary Parameters
FS2
0
0 Programming Manual, 03/2020,
A5E02951525-AB
21