20
Diagnostic Screens
Siemens Data Capture Log
FS200 Utility Version
1.00.02
Long Tag
Product S/N
N1J4118000027
COMS Settings
Baud:19200
Parity:None
Data Bits:8
Location Identifier
Meter Bore id
3.937 inches
System Setup Parameters
------------------------------------------------------------------------------
Parameter
Value
Units
Pipe Diameter
4.537 inches
Pipe Wall Thickness
0.3 inches
Pipe Inside Diameter
3.937 inches
Transducer Size
D4H
______________________________________________________________________________
Path Setup Parameters
Parameter
Path 1
Path 2
Units
Mounting Mode
Direct
Direct
Sensor Spacing LTN
0.4204
0.4204 inches
Sensor Frequency
187500
187500 Hz
Vsmin
197.2924
197.2924 mps
Vsmax
402.7076
402.7076 mps
______________________________________________________________________________
Column#
Label
Description
1
Date
Query Date
2
Time
Query Time
3
Flow Rate(m3/h)
Meter Flow Rate(m3/h)
4
Status FR
Status Flow Rate
7
VF1(m/s)
Flow Velocity path 1(m/s)
9
Avg VF(m/s)
Avg. Flow Velocity(m/s)
10
VoS1(m/s)
VoS path 1(m/s)
12
Avg VoS(m/s)
Avg. VoS(m/s)
13
Status 1
Status path 1
14
Status 2
Status path 2
15
RX Gain(up) 1
Upstream Xmit RX Gain(up) path 1
17
RX Gain(dn) 1
Downstream Xmit RX Gain(dn) path 1
19
DT 1(uS)
Transit-time Delta path 1(uS)
21
SNR(up) 1
Upstream Xmit SNR(up) path 1
23
SNR(dn) 1
Downstream Xmit SNR(dn) path 1
25
CorQ 1
Correlation Factor path 1
27
%ACP 1
Accepted Burst Percentage path 1
29
Temp(F)
Temperature(F)
30
Press(PSIA)
Pressure(PSIA)
31
Density(lbm/ft3)
Density(lbm/ft3)
32
Viscosity(cS)
Viscosity(cS)
Date
Time
Flow Rate(m3/h) Status FR
Standard
Volume Flow
Rate(Nm3/h)
Status SVFR
VF1(m/s)
VF2(m/s)
VoS1(m/s)
3/16/2020
13:19:22
136.1288 Pass
1928.6162 Pass
5.096
4.953 361.49
3/16/2020
13:19:25
136.0898 Pass
1927.9778 Pass
5.126
4.934 361.33
3/16/2020
13:19:28
136.1612 Pass
1928.4122 Pass
5.112
4.948 361.48
3/16/2020
13:19:31
136.0763 Pass
1927.1472 Pass
5.096
4.955 361.52
3/16/2020
13:19:34
135.934 Pass
1924.6659 Pass
5.08
4.976 361.59
3/16/2020
13:19:37
136.0927 Pass
1927.6021 Pass
5.045
4.999 361.49
3/16/2020
13:19:40
136.0052 Pass
1925.5692 Pass
5.03
5.014 361.5
3/16/2020
13:19:43
135.755 Pass
1922.4499 Pass
5.029
4.996 361.45
3/16/2020
13:19:46
135.6573 Pass
1920.7065 Pass
5.058
4.971 361.55
3/16/2020
13:19:49
136.0195 Pass
1927.0039 Pass
5.06
4.984 361.63
3/16/2020
13:19:52
136.1911 Pass
1928.8372 Pass
5.07
4.984 361.61
Average Value
136.0101
1926.2717
5.073
4.974 361.51
Minimum
136.1911
1928.8372
5.126
5.014 361.33
Maximum
135.6573
1920.7065
5.029
4.934 361.63
Warning Limit
Simulation Mode
Simulation Mode
Simulation Mode Simulation Mode > 2%
> 2%
> 0.25%
Failure Limit
Pass
Pass
Pass
Pass
Measurement Measurement > 0.50% Or not Measurement
# ofWarnings
0
0
0
0
0
0
# of Failures
0
0
0
0
0
0
# of Samples
11
Figure 5-6 Capture Log Example
Special Notes for the Data Capture file:
1. The FS200 utility capture file will only contain fields pertinent to the installation. For
example, a Dual Path system with only Path-1 enabled will not present any data columns
FS2
0
0 Programming Manual, 03/2020,
A5E02951525-AB