SIREMOBIL Iso-C 3D
SPR2-230.814.01
Page 2 of 42
Siemens AG
Rev. 05
08.04
CS PS 24
Medical Solutions
6 - 2
3. Part of activities independent of serial number
Replacing the monitor
6
SIMOMED HM
6
•
Remove the monitor on the right as follows.
•
Remove the monitor cover (Fig. 3).
•
Remove the shielding plate (Fig. 4).
•
Remove the connection housing, if in place (Fig. 5).
•
Remove all cables from the monitor to be replaced.
•
Reattach all previously removed parts to the monitor.
•
Remove the 2 screws at the monitor mounting (Fig. 6).
•
Lift down the monitor.
If the TFT displays were ordered and delivered, disregard this
section. Continue with the section "Replacing the monitor, TFT
displays".
Fig. 3
Monitor cover
Fig. 4
Shielding plate
Fig. 5
Connection housing
Fig. 6
Monitor attachment
NOTE
Markierung für Tastaturschublade
Markierung für Tastaturschublade