Route the cables along the guide plate (1/Fig. 29).
•
Route the cable end (2/Fig. 27) to D30 (2/Fig. 29), connect it to connector D30.X11 and
lock it in position. The position of connector D30.X11 is shown in the wiring diagram.
•
Attach the relay (3/Fig. 27) at the location shown in Fig. 30. To do so, remove the existing
screw (1/Fig. 30) and fasten the relay and holding plate using the screw,lock washer and
spacer supplied.
•
Route the cable end (4/Fig. 27) to pc board D200 and connect it to connector
D200.X211.
•
Using the cable ties supplied, fasten the cables along the existing cables. Ensure that the
cables cannot be pinched when the covers are closed.
Fig. 29
Fig. 30
The wiring diagram for the interface cable is attached in
chapter 8.
Please take the wiring diagram out of this document and file it in
the system binder under Wiring Diagram.