SIREMOBIL Iso-C 3D
SPR2-230.814.01
Page 6 of 8
Siemens AG
Rev. 05
08.04
CS PS 24
Medical Solutions
5 - 6
2. Part of activities independent of serial number
Wiring for the orbital drive
5
•
Remove the cover from the encoder (Fig. 19).
•
Fig. 18 shows the plugs from the limit switch cable of the orbital drive.
•
Connect the "Encoder" plug of the encoder cable with the "Encoder" plug under the
cover.
•
Connect the "angulation" plug of the cable for the angulation part with the "X4" plug of the
orbital drive cable set.
•
Connect the "limit switch " plug of the limit switch cable with the "limit switch" plug of the
orbital drive.
Fig. 18 Connectors, cable set for orbital drive
Fig. 19 Plug-in encoder
cover