![Siemens SIMATIC VS130 Series Manual Download Page 70](http://html.mh-extra.com/html/siemens/simatic-vs130-series/simatic-vs130-series_manual_374431070.webp)
Appendix
Vision Sensor VS 130
A-6
A5E00199459-01
Electromagnetic Compatibility (EMC)
Emitted interference
Limit class
•
Radiation of electromagnetic fields complying with EN 55011:
Limit class A, group 1
•
Emission of interference over line a.c. supply complying with
EN 55011: Limit class A, group 1
•
Sensor head and illumination are within the limit value to comply with
EN 55022 class B
Transport and Storage of Modules
With regard to transport and storage conditions, the SIMATIC VS 130 is better than required by
IEC 61131-2. The following information applies to modules transported or stored in their original
packaging.
The climatic conditions correspond to IEC 60721-3-3, Class 3K7 for storage and IEC 60721-3-2,
Class 2K4 for transport.
The mechanical conditions correspond to IEC 60721-3-2, Class 2M2.
Conditions
Permitted Range
Free fall
≤
1 m ( up to 10 kg)
Temperature
-30
°
C to +70
°
C
Air pressure
1080 to 660 hPa (corresponds to a height of -1000 to 3500 m)
Relative humidity (at
+25
°
C)
5 to 95 %, no condensation
Sine-shaped oscillations
complying with IEC 60068-
2-6
5 - 9 Hz:
3.5 mm
9 - 500 Hz:
9.8 m/s
2
Shock complying with IEC
60068-2-29
250 m/s
2
, 6 ms, 1000 shocks
Mechanical Environmental Conditions for Operation
SIMATIC VS 130 is designed for fixed installation in an environment protected from the weather.
SIMATIC VS 130 meets the conditions for use complying with DIN IEC 60721-3-3:
•
Class 3M3 (mechanical requirements)
•
Class 3K3 (climatic environmental conditions)
Mechanical environmental conditions, sine-shaped oscillations
Frequency range in Hz
Test values
10
≤
f < 58
0.075 mm amplitude
58
≤
f < 500
1 g constant acceleration
Test for mechanical environmental conditions
Test for / Test standard
Remarks
Oscillations
Oscillation test complying
with IEC 60068-2-6 (sine)
•
Type of oscillation: Frequency sweeps with a rate of change of 1
octave/minute.
-
10 Hz
≤
f < 58 Hz, constant amplitude 0.075 mm
-
58 Hz
≤
f < 500 Hz, constant acceleration 1 g
-
10 Hz
≤
f
≤
55 Hz, amplitude 1 mm (only sensor head and
illumination)
•
Period of oscillation: 10 frequency sweeps per axis in each of the 3
perpendicular axes