1 - 1
Siemens AG
SPR8-125.061.02
Page 1 of 4
POLYMOBIL Plus
Medical Engineering
Rev. 01
01.00
TD SD 24
Prerequisites
1
Text emphasis
1
Required documents
1
Tools, measurement and auxiliary devices required
1
The tools, measurement and auxiliary devices required, with the exception of the standard
service equipment and the current transformer, are listed in part 3 of the ARTD.
"Warnings" are information provided with special emphasis when
there is the potential for personal injury to the operator or patient.
"Cautions" are information provided with special emphasis when
there is the potential for damage to the equipment.
"Notices" are information provided with special emphasis to facil-
itate proper use of the equipment or proper execution of a proce-
dure.
•
Block circuit diagram
X 069I
•
Function description
SPR8-125.041.01...
A detailed circuit diagram, DIN A 3, can be ordered under item
number 18 20 377 X069I.
•
Standard service kit
•
Digital multimeter
•
2-channel memory oscilloscope
•
mAs meter
•
Protective conductor meter
•
Equivalent leakage current meter
•
Torque wrench 20 Nm - 100 Nm
•
Loctite 242
•
Current transformer (50 A : 50 mA)
with 10 Ohm; 0.5 W; 1% metal film resistor
Refer also to Speed Info 06/91
(31 51 289 B2716)
(10 14 984 B0512)
The oscilloscope has to be connected to ground to perform mea-
surements. A TEK video isolator and the trigger set have to be
used when ground loops may distort the measurement results.
WARNING
CAUTION
NOTICE
NOTE