Line voltage
5
The line voltage must be set between 110 V ( + 10%) or 230 V ( + 10%).
The line voltage is adjusted automatically.
Measuring KVact and Itube
5
•
Connect the oscilloscope to TP "JR", "KV" and "GND" on D915.
•
Set 90 kV and 10 mAs at the POLYMOBIL Plus and release an exposure.
The oscillogram should correspond to the one shown in Fig. 1.
CH1
TP "JR"
1.0 V/Div
CH2
TP "KV"
1.0 V/Div
Ground at "GND"
10 ms/Div
Trigger external at TP "TRG"
Fig. 1