POLYMOBIL Plus
SPR8-125.061.02
Page 6 of 8
Siemens AG
Rev. 01
01.00
TD SD 24
Medical Engineering
2 - 6
Checking the assemblies
Measurement points:
U1 power supply unit 5 V
±
5% / 15 V
±
10 %
2
Fuse: 2 A T
U2 power supply unit 12.5 V
±
10 %
2
Fuse: 5 A T
VP
+15 V
±
10 % for inverter frequency
P12
+ 12.5 V
±
10 % auxiliary voltage
GNDL
Ground (+ 12.5 V)
X4
Pin
Color
1
red
+12.5 V
±
10 %
2
red
+12.5 V
±
10 %
3
red
+12.5 V
±
10 %
4
black
Ground (+12.5 V)
5
black
Ground (+12.5 V)
6
black
Ground (+12.5 V)