
POLYMOBIL Plus
SPR8-125.061.02
Page 2 of 8
Siemens AG
Rev. 01
01.00
TD SD 24
Medical Engineering
6 - 2
Replacing important components
•
Take out one of the four mounting screws and insert it into the threaded hole and tighten
it (arrow/Fig.3).
•
Place a pad under the single tank generator.
•
Disconnect the single tank generator together with the bracket from the stand by
removing the three remaining screws (4/Fig.3).
•
Lower the single tank generator to the pad.
•
The single tank generator is installed in reverse sequence.
- Secure the three mounting screws (4/Fig.3) with Loctite 242 and tighten them.
- Remove the fourth mounting screw from the threaded hole (arrow/Fig.3) and reinstall it
in its former position. Secure the screw with Loctite 242.
- Test the torque of the mounting screws ;
Nom: 25 Nm, Tolerance
±
10%.
•
Reinstall the collimator in reverse sequence.
•
Set the tube current (refer to "Setting the tube current for the exposure", pp 5-12).
•
Check the coincidence of the light and radiation fields and adjust, if necessary
(refer to "Coincidence of light and radiation fields", pp 5-15)
This is a safety device to prevent the arm of the unit from flying
upward if the lever for the transport lock is inadvertantly
activated.
4
Fig. 3
WARNING