Measurements
5 - 5
Siemens AG
SPR8-125.061.02
Page 5 of 20
POLYMOBIL Plus
Medical Engineering
Rev. 01
01.00
TD SD 24
Rotating anode starter
5
Acceleration
5
Fig. 5 shows the stator currents during rotating anode acceleration (230 V).
The currents are correspondingly lower at 100V line voltage.
Fig. 5
CH2 D915 test point "I1" (1 V 0.25 A)
CH1 D915 test point "I2" (1 V 0.25 A)
10 ms /Div
NOTICE