5 - 4
Measurements
Capacitor voltage (Uc) and charging time
5
In exposure-ready status, capacitor voltage is approximately 440 V and can be measured
on D 950 TP "VC".
If the device is switched off because of overvoltage, the signal "ERR" is output on D950.
The charging time depends upon the residual charge of the capacitor bank and on the
available line voltage, e.g. with 230 V the max. charging time is about 15 s.
In the worst case, ( line voltage <100 V and residual charge = 0V), charging time
can take 40 seconds.
For the following measurement, the line voltage was 230 V and the residual charge 30 V.
You can measure the capacitor voltage at TP "VC" only if
POLYMOBIL Plus is switched on.
CAUTION
Fig. 4
D915 test point "VC" (1 V 100 V)