Page 10
Pointek CLS 300 – INSTRUCTION MANUAL
7ML19985CK03
Rod Version
The rod version of the CLS 300 is available in standard lengths from 350mm (14") up to 1000mm
(40"). Other lengths using stainless steel or other metals can be attached as an electrode
when required. Ensure that the thread end conforms to the exact dimensions as shown
below. Rod lengths beyond 1000mm (40") are not recommended.
Changing Probe Length
The probe length can be shortened in the field by cutting the electrode.
Warning:
To prevent damage, do not apply torque directly to the main probe assembly.
Option 1
If the electrode is removable:
1.
Remove the electrode by
releasing the set screw and
turning the threaded electrode
end counter-clockwise.
2.
Place the upper part of the
electrode in a vice as shown.
3.
Use a wrench to loosen the
lower portion of the electrode.
12mm (0.47")
3 mm (0.12”)
1.5mm by 45
(0.06")
Ø19 mm
(0.75" )
Ø11 mm
(0.43”)
M5
9mm
(0.35")
17mm
(0.67")
34mm
(1.3") 37mm
(1.5")
M10
Flat sides
= 18mm
(0.71")