4-6
February 5, 2004
FFDM-S Service Manual
2. Once the setup is complete, click the
Enable Capture
button.
The Status message changes from “DRAPI Ready” to “DRAPI Armed” and
turns from yellow to green.
3. Make sure the Generator is configured properly, then make the indicated
exposure.
4. Select the desired units.
5. In the
Reading
field, enter the dosimeter value in the appropriate units: mR
(milliRoentgen) or mGY (milliGray).
6. Make sure the entry is correct, then click the
Submit
button.
As the DirectRay Detector and DRAPI recycle for the next exposure, the Status
will change from “DRAPI Armed” to “DRAPI Not Ready” and will turn from
green to white.
When the DRAPI is ready to be enabled for the next exposure, the Status will
again say “DRAPI Ready” and will be yellow. Instructions for the next exposure
display.
7. Perform the displayed instructions.
The amount of aluminum added in this step is dependent on the target/filter
selection (Mo/Mo = 0.3mm; Mo/Rh = 0.3mm; W/Rh = 0.5mm).
8. Click the
Enable Capture
button.
The last dosimeter reading you entered is cleared. The status window will
change from “DRAPI Ready” to “DRAPI Armed” and will turn from yellow to
green.
Summary of Contents for MAMMOMAT NovationDR
Page 7: ...vi February 5 2004 FFDM S Service Manual THIS PAGE IS INTENTIONALLY BLANK ...
Page 29: ...2 12 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Page 47: ...4 10 February 5 2004 FFDM S Service Manual ...
Page 51: ...4 14 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Page 68: ...PXCM Web Tool February 5 2004 5 17 ...
Page 76: ...PXCM Web Tool February 5 2004 5 25 ...
Page 83: ...6 4 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Page 95: ...B 2 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...