
3-6
February 5, 2004
FFDM-S Service Manual
DirectRay PXCM Installation
Mount the DirectRay PXCM in the mammography equipment on four size M4
studs, in the position indicated in Figure 3-3.
NOTE:
To ensure reliable communications, keep the fiber optic connections covered by protective
caps when not in use.
Figure 3-3.
DirectRay PXCM Stud Hole Locations
Figure 3-4 shows the PXCM connectors used for the cable that connects the
Detector to the DirectRay PXCM, and the fiber optic cable to the DirectRay
Interface card.
Figure 3-4.
DirectRay PXCM Connectors (for cable to Detector and fiber optic cables to
Interface card)
GS20_412.cdr
Connector for cable
to DirectRay Detector
Fiber optic connectors
to DirectRay Interface card
Connector for cable
to Generator
Connector for PXCM
power cord
PXCM power switch
Summary of Contents for MAMMOMAT NovationDR
Page 7: ...vi February 5 2004 FFDM S Service Manual THIS PAGE IS INTENTIONALLY BLANK ...
Page 29: ...2 12 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Page 47: ...4 10 February 5 2004 FFDM S Service Manual ...
Page 51: ...4 14 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Page 68: ...PXCM Web Tool February 5 2004 5 17 ...
Page 76: ...PXCM Web Tool February 5 2004 5 25 ...
Page 83: ...6 4 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Page 95: ...B 2 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...