1-8
February 5, 2004
FFDM-S Service Manual
CAUTION
Some of the electronic components used in this product may be damaged
by static electricity or electrostatic discharge. Unpack and handle using appropriate
precautions.
CAUTION
Dielectric strength test caution: (Statement from Condor DC Supplies Inc.
regarding their GSM28-5 power supply) "When performing Dielectric Strength Tests,
catastrophic failure of the unit may result if a Dielectric Strength test voltage greater
than 1800 Vac is applied between primary and secondary circuits. The components
providing isolation from primary to secondary cannot be tested while installed in the
power supply without overstressing basic (primary to ground) insulation. All isolating
components are individually 100% tested at 4800 Vac prior to installation."
Summary of Contents for MAMMOMAT NovationDR
Page 7: ...vi February 5 2004 FFDM S Service Manual THIS PAGE IS INTENTIONALLY BLANK ...
Page 29: ...2 12 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Page 47: ...4 10 February 5 2004 FFDM S Service Manual ...
Page 51: ...4 14 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Page 68: ...PXCM Web Tool February 5 2004 5 17 ...
Page 76: ...PXCM Web Tool February 5 2004 5 25 ...
Page 83: ...6 4 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Page 95: ...B 2 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...