Maintenance and Service
6.7 Reconfiguration of the Path Length
LDS 6
Operating Instructions, 01/2009, A5E00295894-05
75
6.7
Reconfiguration of the Path Length
1.
Starting from "Measuring window", press the soft key next to the component for which
you want to change the path length. Press the soft key for "Configuration". Enter the
password for privilege level 2 (the password is factory-set to the value "222", but may
have been replaced by a new one). Press "Continue" twice and then select function 84
"Path Length".
2.
Press first soft key to edit the path length. Enter the new path length and press ENTER.
3.
Press MEAS to return to measurement screen. Press MEAS again to loose privileges.
4.
The procedure has to be repeated for other channels. It is not necessary to repeat the
procedure for more than one component per channel since this function is a channel-
specific function.