Operation
4.3 Analyzer Functions
LDS 6
58
Operating Instructions, 01/2009, A5E00295894-05
74 Reset
74 Reset Ch1 NH
3
Trigger Reset
You can use this function to carry out a cold restart of the analyzer, i.e. in case of a fault in
program execution.
You must wait for the start-up time to elapse, before trying to use the display. The analyzer will
automatically start to measure and will be ready for use after 1 to 3 minutes.
75 Save/Load Data
75 Save/Load Data Ch1 NH
3
Load Factory Settings
Load User Data
Save User Data
Erase EEProm Section
You can use this function to save or load user-specific data in the user data memory on the
EEPROM. Saving of data should always be carried out after a successful starting-up of the
system. All individual settings are then saved and can be recalled if necessary (load user data).
This is significant if repairs or maintenance are to be carried out on an analyzer or if new
parameter settings are to be tried.
75 Erase Section Ch1 NH
3
Erase User Section
Erase Work Section
You can use this function to erase the working area data and the user data on the EEPROM.
The factory data can never be erased.