Operation
4.3 Analyzer Functions
LDS 6
Operating Instructions, 01/2009, A5E00295894-05
49
4.3.2
Analyzer Status
Table 4- 3
4 Display Meas. Range
3 Logbook
2 Diagnostics Values
1 Analyzer Configuration
Analyzer Status Ch1 NH
3
This display appears when pressing the first soft key ("Analyzer Status"), following the
selection of the diagnostics functions in the main menu.
The status functions are freely-accessible. You will not be asked to enter a password.
1 Analyzer Configuration
1 Analyzer Config. Ch1 NH
3
CE Drv.Software
1 .0 .314
CE Software
3 .0 .126
OS Version
60001
Serial No.
7MBXXXXXXXXXXXXXXXXX
Order No.
...Continue
X
2 .3 .245
23
Software Package Version
You can use this function to display important manufacturing data of the analyzer.
● Ordering No: Information on ordering data of analyzer.
● Serial No: Consecutive number of analyzer.
● Software Package Version: Version number of the application-specific configuration
software.
● OS Version: Version number of the Windows CE operative system running in the
analyzer.
● CE Software: Version number of the LDS 6-specific software.
● CE Driver Software: Version number of the driver software.
● Analyzer uC Drv.Software: Version number of the software running on the main micro
controller.
● Channel uC Drv.Software: Version number of the software running on the measurement
channel micro controller.
● Opto uC Drv.Software: Version number of the software running on the opto module micro
controller.
● Chan. FPGA Firmware: Version number of the firmware running on the measurement
channel FPGA.
● Opto. FPGA Firmware: Version number of the firmware running on the laser controller
FPGA.
● Ref. FPGA Firmware: Version number of the firmware running on the reference channel
FPGA.