CR-IR347
Service Manual
PM - 80
009-058-00
10.20.2000 FM2732 (2)
PM - 80
9.7
Subscanning Unit (Cleaning the Light-Collecting Guide
Assemblies)
1Y
At intervals of one year, or when 60,000 IPs have been processed
→
Clean.
◆
NOTES
◆
• Wipe the light-collecting guide assembly in one direction at a constant rate.
• If the lens cleaning paper becomes soiled, replace it with a new one.
• Do not wipe the photomultiplier bonding surface.
Light-collecting guide assembly
(for the front)
Light-incident face
Light-collecting guide assembly
(for the back)
Light-incident face
Photomultiplier
bonding surface
Socket
section
PMT08D/
PMR08D board
Cleaning range
5 mm approx.
FR7H6062.EPS
Light-collecting guide assembly
[Clean]
Lens cleaning paper (ethanol)
1Y
Summary of Contents for CR-IR 347
Page 2: ...CR IR347 CR IR347P Service Manual Preventive Maintenance PM ...
Page 22: ...CR IR347 Service Manual PM 15 3 009 058 03 08 30 2002 FM3476 PM 15 3 BLANK PAGE ...
Page 23: ...CR IR347 Service Manual PM 15 4 009 058 03 08 30 2002 FM3476 PM 15 4 BLANK PAGE ...
Page 29: ...CR IR347 Service Manual PM 21 009 058 00 10 20 2000 FM2732 2 PM 21 BLANK PAGE ...
Page 59: ...CR IR347 Service Manual PM 51 009 058 00 10 20 2000 FM2732 2 PM 51 BLANK PAGE ...
Page 73: ...CR IR347 Service Manual PM 65 009 058 00 10 20 2000 FM2732 2 PM 65 BLANK PAGE ...
Page 83: ...CR IR347 Service Manual PM 75 009 058 00 10 20 2000 FM2732 2 PM 75 BLANK PAGE ...
Page 89: ...CR IR347 Service Manual PM 81 009 058 00 10 20 2000 FM2732 2 PM 81 BLANK PAGE ...
Page 110: ...CR IR347 Service Manual PM 102 009 058 00 10 20 2000 FM2732 2 PM 102 BLANK PAGE ...
Page 119: ...CR IR347 Service Manual PM 111 009 058 00 10 20 2000 FM2732 2 PM 111 BLANK PAGE ...