CR-IR347
Service Manual
PM - 52
009-058-00
10.20.2000 FM2732 (2)
PM - 52
8.5
Erasure Conveyor (Cleaning/Replacing the Shock
Absorber-Attached Filter)
1Y
At intervals of one year, or when 60,000 IPs have been processed
→
Clean
3Y
At intervals of three years, or when 180,000 IPs have been processed
→
Replace
◆
NOTE
◆
When cleaning the shock absorber-attached filter with a moistened cloth, use care not to touch
the shock absorber.
FR7H6043.EPS
T
O
P
Shock absorber-attached filter
Shock absorber
(1) Remove the lamp house assembly.
Erasure conveyor
FR7H6171.EPS
[Remove] BR4x8 (x4)
Lamp house assembly
Summary of Contents for CR-IR 347
Page 2: ...CR IR347 CR IR347P Service Manual Preventive Maintenance PM ...
Page 22: ...CR IR347 Service Manual PM 15 3 009 058 03 08 30 2002 FM3476 PM 15 3 BLANK PAGE ...
Page 23: ...CR IR347 Service Manual PM 15 4 009 058 03 08 30 2002 FM3476 PM 15 4 BLANK PAGE ...
Page 29: ...CR IR347 Service Manual PM 21 009 058 00 10 20 2000 FM2732 2 PM 21 BLANK PAGE ...
Page 59: ...CR IR347 Service Manual PM 51 009 058 00 10 20 2000 FM2732 2 PM 51 BLANK PAGE ...
Page 73: ...CR IR347 Service Manual PM 65 009 058 00 10 20 2000 FM2732 2 PM 65 BLANK PAGE ...
Page 83: ...CR IR347 Service Manual PM 75 009 058 00 10 20 2000 FM2732 2 PM 75 BLANK PAGE ...
Page 89: ...CR IR347 Service Manual PM 81 009 058 00 10 20 2000 FM2732 2 PM 81 BLANK PAGE ...
Page 110: ...CR IR347 Service Manual PM 102 009 058 00 10 20 2000 FM2732 2 PM 102 BLANK PAGE ...
Page 119: ...CR IR347 Service Manual PM 111 009 058 00 10 20 2000 FM2732 2 PM 111 BLANK PAGE ...