Environmental Specifications
Electrical Requirements
Specification
Value
Ambient Operating
Temperature
18°C to 30°C (64°F to 86°F).
At temperatures under 22°C (72°F), urobilinogen
and leukocyte results may be decreased, and at
temperatures above 26°C (79°F), increased.
Optimum Operating
Temperature
22°C to 26°C (72°F to 79°F).
Relative Humidity
20% to 80%, non-condensing, actively controlled
Optimum Relative
Humidity
35% to 55%
Indoor Use Only
Altitude
up to 2000 meters
Ventilation
1709 BTU
IEC 1010-1 Installation Category II
IEC 1010-1 Equipment Classification Class I
IEC 1010-1 Pollution Degree 2
Requirement
Value
Electrical Rating
100 to 240 VAC
±
10%
Power Requirements
50 to 60 Hz
Maximum Power Input
72 VA
Fuse Rating
2 A, 250 V, 2 AG, SB(T)
Line Leakage Current
<0.5 mA in normal condition
<3.5 mA in single fault condition