Error 30
Error 31
Error 34
Analyzer
mechanical error
Contact your local technical support
provider.
Error 36
Both areas of
analyzer memory
where factory
calibration
parameters are
stored are corrupt
1. Turn analyzer power off.
2. Wait several seconds.
3. Turn analyzer power on.
4. If the error repeats, contact your
local technical support provider.
Error 37
Touch screen
calibration error
1. Follow the instructions on the
display to calibrate the touch
screen.
If the error repeats:
1. Turn analyzer power off.
2. Contact your local technical
support provider.
Error 50
Printer Error
1. Ensure that the external printer
is turned on and is online.
2. Verify that both ends of the
interface cable are securely
connected.
3. Check that the printer has
paper.
Error 51
Error 52
Quality control
results memory
(51) or sample
results memory
(52) is almost full
Nearly 200 quality control result
sets or nearly 500 patient result
sets are stored in memory and have
not been transferred to a computer.
1. Ensure that the computer is
turned on.
2. Ensure that the interface cable
is securely connected at both
ends
3. Ensure that the setup
parameters for the computer
interface are correct.
4. Transfer at least some of the
records.
5. If unable to transfer records,
contact your local technical
support provider.
Symptom
Possible Cause
Remedy