Error 23
Moving table is
misaligned
1. Turn analyzer power off.
2. Remove the push bar, the fixed
platform, and the moving table.
You may have to pull firmly to
remove these items.
3. Turn analyzer power on.
4. Allow the analyzer to reinitialize,
the table mechanism to move to
its lowest position, and another
error displays.
5. Turn analyzer power off.
6. Reinstall the moving table.
7. Ensure it is pushed in
completely.
8. Reinstall the fixed platform, and
push bar.
9. Turn analyzer power on.
Analyzer
mechanical error
Contact your local technical support
provider.
Error 24
Error 25
Fixed platform is
misaligned or
push bar is
misaligned or
upside down
1. Turn analyzer power off.
2. Inspect the analyzer for any
obvious signs of misalignment
or incorrect installation of the
push bar, the fixed platform, or
the holddown plate.
3. Remove and reinstall, if needed.
4. Ensure the feet on the push bar
are on the bottom, nearest the
platform.
5. Turn analyzer power on.
Analyzer
mechanical error
Contact your local technical support
provider.
Symptom
Possible Cause
Remedy