o
n
Color
Use the
Color
cycle key to have the analyzer determine color, or allow visual
determination.
NOTE:
The analyzer can only determine color if the Siemens Diagnostics
Reagent Strip used contains the leukocyte test. Results reported by the
analyzer may be different from the color seen visually. This is because of the
inherent differences between the human eye and the optical system of the
analyzer.
Color Choices
If the color option is Entered by tech, you can specify up to 7 specimen colors.
Use this procedure to customize descriptions and remove defaults from the
reporting list.
1. Select
Color choices
.
2. Edit the first 4 default colors.
3. Select
Next Screen
to display and edit the last 3 default colors.
The default colors are Yellow, Orange, Red, Green, Blue, Brown, and
Other.
4. Remove default options from the reporting list.
The colors included on the list are designated by a check mark.
a. Select the check mark to remove it, and delete the option from the list.
The first option is always selected and cannot be made inactive.
5. Change the color description:
a. Select the word describing the color to change the color.
An alphabetic keypad displays.
b. Use the
Move Left
key to erase the existing name.
c. Enter the new name.
You can use up to 15 letters and spaces.
d. Select
Enter
.
6. Select
Previous Screen
.
To...
Select...
have the analyzer determine color
automatically,
Determined by analyzer
enter the color as part of a manually
entered loadlist or just before testing
each specimen,
Entered by tech