3. If necessary, select
Next Screen
to display an additional screen of tests.
4. When finished, select
Previous Screen
.
Normal Range for SG and pH
You can select this option only if Mark positives is On.
Use this procedure to set the lower and upper limits of the normal range for SG
and pH. Set each limit separately. The upper limit must be higher than or equal
to the lower limit.
1. Select
Normal range for SG/pH
.
2. Select the
+
or
-
keys next to each limit to raise or lower the limit.
The limit changes by 1 reporting level until it is equal to the opposite limit or
is at the highest or lowest reporting level.
3. Select
Previous Screen
.
Normal Range for CRE
You can select this option only if you are using MULTISTIX PRO Reagent
Strips and Mark positives is On.
Use this procedure to select the lower and upper limits of the normal range for
creatinine. The upper limit must be equal to or higher than the lower limit.
1. Select
Normal range for CRE
.
2. Select the
+
or
-
keys next to each limit to raise or lower the limit.
The number changes by 1 reporting level until it is equal to the opposite
limit or is at the highest or lowest reporting level.
3. Select
Previous Screen
.
Setup Menu 5
At the fourth Setup menu, select
Next Screen
to access the fifth Setup menu.
Menu
Option
Default
Color
Determined by analyzer
Color choices
Yellow, Orange, Red, Green, Blue,
Brown, Other
Clarity choices
Clear, SL Cloudy, Cloudy, Turbid,
Other
Use default COL/CLA during run
On