Operation
5-46
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
You can carry out the following tests:
S
Test analog outputs
You can use this dialog to set any value of the output current
between 0 and 20 mA. There are four analog outputs (0/4 to
20 mA). To test these outputs, connect an ammeter to the
corresponding analog outputs on plug X80 and measure the output
current.
Please refer to Chapter 3 for the pin assignments of plug X80.
S
Test of relays
You can use this dialog to select a relay for testing. There are eight
relays whose contacts you can test. Connect a ohmmeter to the
corresponding relay outputs on plug X80. You can process the
following parameters:
Relay number:
One of the relays 1 to 8 (for analyzers with option
module: relays 1 to 16) or ”-” (no relay)
Condition:
The current state of the selected relay (
active
or
inactive
; inactive in the Fig.).
Please refer to Chapter 3 for the pin assignments of plug X80/X50.
S
External solenoid valve
You can use this dialog to trigger external solenoid valves for the
zero and calibration gas supplies via the relay contacts. Use the
arrows keys
↑
and
↓
to select either the zero gas valve (
ZV
) in the
first line and the calibration (span) gas valve (
SV
) in the third line.
Call the selected valve using the
ENTER
key. Switch the previously
assigned relay using any arrow key in the second or fourth line (the
value on the right toggles between
OFF
and
ON
). The measured
values output on the display during this test usually change following
the switching operation.
S
Test of inputs (binary test)
After calling this dialog, apply a voltage of 24 V to one of the three
inputs ”Pump ON/OFF”, ”CAL” or ”SYNC”. The result is displayed in
the fourth line (no input in this case).
The external binary inputs of analyzers with option module can be
tested in the menu ”Binary inputs” (see Section 5.10.1.3).
Analog 1: 0.20 mA
Analog 2: 0.40 mA
Analog 3: 1.55 mA
Analog 4: 2.60 mA
51032a
Test
Test of relays
Relay number 1
Condition: inactive
Press ESC to cancel
51032b
51032c
130 mg/m
3
NO
ZV
89 mg/m
3
CO
OFF
249 mg/m
3
SO
2
SV
20.77 %
O
2
OFF
Test of inputs
Press ESC to cancel
Active input:
None
51032d
Summary of Contents for 7MB2335
Page 2: ......
Page 8: ...I 4 Operating Instructions ULTRAMAT 23 Gas Analyzer C79000 B5276 C216 03 ...
Page 112: ...Shut down 7 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 113: ...8 1 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Spare Parts List 8 ...
Page 133: ...Spare parts list 8 21 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 137: ...Spare parts list 8 25 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 164: ...Appendix 9 12 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 166: ...Appendix 9 14 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 167: ......