Operation
5-21
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
5.8
Calibration
Using this function it is possible to calibrate the IR channels of the
ULTRAMAT 23 using one or more calibration gases, and to reset the
sensitivity. In addition, you can calibrate the oxygen sensor and the
pressure sensor and also define the AUTOCAL parameters. Fig. 5-5
shows the associated menu sequence (see Section 5.7 for description
of display elements).
Calibration is only possible if you have enabled code level 1.
5.8.1
Calibration: Calibrate IR Channels
In this menu you can:
S
Set, in line 2, the setpoints of the calibration gases for the individual
ranges or select a total or single calibration.
S
Start a calibration procedure in lines 3 and 4.
This function is specific to the component.
5.8.1.1
Calibration: Calibrate IR Channels: Set Span Gas Values
S
Total cal.:
In the first line of this dialog, you can select either a total
or single calibration.
ON
means that one range is calibrated and that this calibration is
used for the other ranges (total calibration).
OFF
means that each range is calibrated separately (single
calibration, e.g. with different calibration gases).
S
With AUTOCAL
: You can define here whether you wish to carry out
an AUTOCAL prior to the calibration procedure (
YES
or
NO
). An
AUTOCAL is not necessary here if it has already been carried out
shortly before the calibration procedure, e.g. a calibration procedure
has taken place directly previously.
S
Span1, Span2
: Here you can enter the setpoints for the individual
measuring ranges (these are the concentrations of the components
in the respective calibration gas). They should be set to a value
which is between 70 and 100 % of the full-scale value. If
Total cal.:
ON
has been selected, the analyzer automatically uses the setpoint
of range 2 for range 1. With
Total cal.: OFF
any input is possible
between the start-of-scale and full-scale values of the respective
range.
Calibration SO
2
Set span gas values
Start with range 1
Start with range 2
581a
Total cal.: OFF SO
2
With AUTOCAL: NO
Span 1 :
386 mg/m
3
Span 2 : 1920 mg/m
3
5811
Total cal.: OFF SO
2
With AUTOCAL: NO
Span 1+2 :
12 %
Span 1+2 :
12 %
5811a
Summary of Contents for 7MB2335
Page 2: ......
Page 8: ...I 4 Operating Instructions ULTRAMAT 23 Gas Analyzer C79000 B5276 C216 03 ...
Page 112: ...Shut down 7 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 113: ...8 1 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Spare Parts List 8 ...
Page 133: ...Spare parts list 8 21 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 137: ...Spare parts list 8 25 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 164: ...Appendix 9 12 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 166: ...Appendix 9 14 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 167: ......