Operation
5-12
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
5.5.3
Using the ESC key
You can trigger two different functions by pressing the
ESC
key:
Firstly, you can cancel a commenced procedure, e.g.:
S
The input of a number
S
A calibration procedure with calibration gas
S
Any function if a fault occurs, e.g. if the flow of sample gas to the
analyzer is missing.
Secondly, you can use the
ESC
key to move one level higher in
submenus (”scroll back”). This procedure is the opposite to selection of
a submenu using the
ENTER
key (”scroll forwards”). If you repeatedly
press the
ESC
key, you return back to the main menu step-by-step. If
you press the
ESC
key again in the main menu, the analyzer switches
over from input to measuring mode. All inputs are imported at the same
time. However, you will not be asked to confirm the inputs.
An example will clarify this:
The analyzer is in measuring mode and is uncoded.
Switch from measuring mode to input mode using
ENTER
, select the
menu item ”Parameters” using the
↑
or
↓
key, and press
ENTER
.
In this manner, you enter the first submenu. Press
ESC
and then
ENTER
again. You have returned by one level and then moved
forwards by one level again; you are therefore in the same menu again.
Press the
ESC
key twice, you are back in measuring mode again.
403.8
vpm CO
|
1875
vpm NO
|
936.4
vpm SO
2
|
4.2
%
O
2
| U
Analyzer status
Calibration
P
arameters
Configuration
M
easuring ranges
Limit values
Time constants
Pump/LCD contrast
403.8 vpm
CO
|
1875
vpm
NO
|
936.4 vpm
SO
2
|
4.2
%
O
2
| U
Summary of Contents for 7MB2335
Page 2: ......
Page 8: ...I 4 Operating Instructions ULTRAMAT 23 Gas Analyzer C79000 B5276 C216 03 ...
Page 112: ...Shut down 7 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 113: ...8 1 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Spare Parts List 8 ...
Page 133: ...Spare parts list 8 21 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 137: ...Spare parts list 8 25 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 164: ...Appendix 9 12 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 166: ...Appendix 9 14 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 167: ......