Operation
5-36
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
5.10.1.2
Configuration: Inputs/Outputs/Pump: Assign Relays
You can use this menu to assign different functions to up to eight
relays which are installed in the analyzer. These functions can be
signals for limits or ranges, status signals, and the functions of the
external solenoid valves. If an option module is present, eight additional
relays, i. e. a total of 16 relays, can be assigned to the corresponding
functions.
Each function may only be assigned to one single relay. The analyzer
outputs an error message if you attempt a double assignment for a
relay which has already been assigned. A relay to which a function has
not been assigned is shown on the display by a dash.
Table 5-5
Summary of relay assignments
Function
Relay de-energized
Relay energized
Signalling
Limit
Limit has been triggered
Limit, see Section 5.9.2
Measuring range
Range 2
Range 1
Fault
Fault present
Maintenance request
Maintenance request
CAL/MEAS
Measure
AUTOCAL
AUTOCAL
Function control
Function control present
With warm-up mode
(approx. 30 min),
AUTOCAL, uncoded
Zero gas
Zero gas flowing
Ext. solenoid valve open
Calibration gas
(Sample gas inlet)
Calibration gas flowing
Ext. solenoid valve open
Sync.
Synchronization signal
is output
AUTOCAL only ”Zero
gas flow” and
adjustment not during
sample gas purge
phase
The pin assignments of the relays in the de-energized state are described in Chapter 3.
S
Relays for limits
The upper and lower limits can be defined as events for triggering
relays. Select the desired relay number in the second and third
lines.
This function is specific to the component.
Relays for limits
Relays for MR
Relays for status
Relays for valves
51012
Relays/limits SO
2
Limit 1: Relay 1
Limit 2: Relay 2
51012a
Summary of Contents for 7MB2335
Page 2: ......
Page 8: ...I 4 Operating Instructions ULTRAMAT 23 Gas Analyzer C79000 B5276 C216 03 ...
Page 112: ...Shut down 7 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 113: ...8 1 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Spare Parts List 8 ...
Page 133: ...Spare parts list 8 21 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 137: ...Spare parts list 8 25 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 164: ...Appendix 9 12 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 166: ...Appendix 9 14 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 167: ......