Operation
5-18
Operating instructions ULTRAMAT 23 gas analyzer
C79000-B5276-C216-03
5.7.2.2
Analyzer Status: Diagnostic Values: O
2
Diagnostic Values
This dialog displays the diagnostic values of the optional oxygen
sensor.
S
Sensorsig
is the current pessure-compensated voltage of the O
2
sensor in mV
S
O
2
is the current oxygen value (negative values are also possible)
S
Date
is the installation date of the O
2
sensor (see Section 5.8.2)
S
Inst.volt
is the pressure-compensated voltage of the O
2
sensor
when it was installed.
5.7.2.3
Analyzer Status: Diagnostic Values: Pressure Sensor Diagnosis
This dialog displays the diagnostic values of the pressure sensor (see
Section 5.8.3). The displayed values have the following meaning:
S
ADU pressure
is the voltage of the pressure sensor measured at
the output of the A/D converter.
S
Pressure air
is the current atmospheric pressure in mbar.
5.7.2.4
Analyzer Status: Diagnostic Values: Other Diagnostic Values
This menu is used to call further diagnostic functions. You can call the
following dialogs:
S
Power/IR source
Power:
Data on the mains voltage in % of the nominal value of the
respective mains voltage (e.g.: 100 % corresponds to 230 V or
120 V).
Raw voltage:
Raw voltage following the rectification
Source 1, Source 2:
Data on the IR source voltage(s), in Volt
provided the IR source is present.
S
Reference/Bridge
Reference:
The reference voltage for the electronics of the
analyzer.
Bridge:
The supply voltage to the measuring bridge.
S
LCD temperature
The temperature which determines the display contrast. Refer to
Section 5.9.4.2 for adjustment of the LCD contrast.
Sensorsig:
12.02 mV
O
2
:
20.94 %
Date: 27.05.1996
Inst.volt.: 12.10 mV
5722
Pressure ADU: 3.965 V
Pressure air: 999 mbar
5723
Power/IR source
Reference/Bridge
Temperature of LCD
Analog output
5724
Power (VAC) :
101 %
Raw voltage :
30 V
Source 1
: 7.541 V
Source 2
:15.023 V
5724a
Reference
:
2.513 V
Bridge
:
3.755 V
5724b
LCD temp.
: 34.9
°
C
5724c
Summary of Contents for 7MB2335
Page 2: ......
Page 8: ...I 4 Operating Instructions ULTRAMAT 23 Gas Analyzer C79000 B5276 C216 03 ...
Page 112: ...Shut down 7 4 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 113: ...8 1 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 Spare Parts List 8 ...
Page 133: ...Spare parts list 8 21 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 137: ...Spare parts list 8 25 Operating instructions ULTRAMAT 23 Gas analyzer C79000 B5276 C216 03 ...
Page 164: ...Appendix 9 12 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 166: ...Appendix 9 14 Operating instructions ULTRAMAT 23 gas analyzer C79000 B5276 C216 03 ...
Page 167: ......