![Siemens Healthcare Clinitek Status Connect System Operator'S Manual Download Page 27](http://html.mh-extra.com/html/siemens-healthcare/clinitek-status-connect-system/clinitek-status-connect-system_operators-manual_1266125027.webp)
Clinitek Status Connect System
27
Operator’s Guide
5. To permit patient tests if the QC test fails, select
Yes
.
To prevent patient tests if the QC test fails, select
No
.
6. Select
Next
.
The
QC Settings-Strip test
screen 3 of 3 displays.
NOTE:
If you selected to have the operator determine pass or fail in Step 4
above, the
Defining QC Strip Test Controls
and
Setting QC Strip Test
Pass/Fail Ranges
below, are not necessary.
NOTE:
If you selected to have the operator determine pass or fail in Step 4
above, you must use the strip selected in the Urinalysis Test Settings screen,
a color-banded strip, or an IR-banded strip. You can only run one QC level for
each QC test; additional levels require an additional QC test.
Defining QC Strip Test Controls
To define QC strip test controls, perform the following steps:
1. To set the number of levels to complete a QC test, select
1
,
2
, or
3
.
2. For each control set in Step 1, select the corresponding
Control Level, 1
,
2
, or
3
.
For example, if you set 2 controls, you must define Control Level 1 and
Control Level 2.
The
Control Level 1-Strip test
screen displays.
3. Select
Enter name of control
.
The
Name of Control
screen displays.
4. Enter the name of this control.
Use the alpha keyboard to enter text.
To enter numeric text, select
123
.
5. Select
Enter
.
The
Control Level 1-Strip test
screen displays.
6. Select
Enter name of level
.
The
Name of Level
screen displays.
7. Enter the name of this level.
Use the alpha keyboard to enter text.
To enter numeric text, select
123
.
8. Select
Enter
.
The
Control Level 1-Strip test
screen displays.
9. Select
Select strip type
.
The
QC Strip Type-Select strip for this QC level
screen displays.
10. Use the arrow keys to select the strip type.
11. Select
Select
.
The
Control Level 1-Strip test
screen displays.
Summary of Contents for Clinitek Status Connect System
Page 1: ...Clinitek Status Connect System Operator s Guide For US Only 10490852 2009 12 135055 Rev A ...
Page 6: ...Clinitek Status Connect System 6 Operator s Guide ...
Page 14: ...Clinitek Status Connect System 14 Operator s Guide ...
Page 24: ...Clinitek Status Connect System 24 Operator s Guide ...
Page 36: ...Clinitek Status Connect System 36 Operator s Guide ...
Page 38: ...Clinitek Status Connect System 38 Operator s Guide ...
Page 54: ...Clinitek Status Connect System 54 Operator s Guide ...
Page 56: ...Addresses www siemens com diagnostics Clinitek Status Connect System 56 Operator s Guide ...
Page 58: ...Clinitek Status Connect System 58 Operator s Guide ...
Page 64: ...Clinitek Status Connect System 64 Operator s Guide ...
Page 68: ...Clinitek Status Connect System 68 Operator s Guide ...
Page 72: ...Clinitek Status Connect System 72 Operator s Guide ...